Our Products

Product Range

product_image5824
product_image5824im3536_2__0
Hioki's LCR Meters and Impedance Analyzers cover a frequency range from 1 MHz to 3 GHz, catering to diverse applications in electronic component testing. The IM3536 sets a new benchmark for general-purpose LCR meters, offering a broad testing frequency range from DC and 4 Hz to 8 MHz (*1) with an im ...Read more
HIOKI - IM3536 LCR Meter

 FEATURES: 

  • Wide Measurement Frequency Range: DC, 4 Hz to 8 MHz (customizable up to 10 MHz by special order).

  • High-Speed Measurements: Achieves measurement speeds as fast as 1 ms.

  • High Precision: Delivers measurement accuracy of ±0.05% (representative value).

  • Low Impedance Measurement: Guaranteed accuracy starting from 1 mΩ, ensuring reliable low-impedance measurements.

  • DC Bias Function: Allows measurements under conditions that simulate actual use or comply with industry standards.

  • Versatile Applications: Suitable for a broad spectrum of uses, from research and development to production lines.

 

 




  • Details
  • Specifications

Hioki's LCR Meters and Impedance Analyzers cover a frequency range from 1 MHz to 3 GHz, catering to diverse applications in electronic component testing. The IM3536 sets a new benchmark for general-purpose LCR meters, offering a broad testing frequency range from DC and 4 Hz to 8 MHz (*1) with an impressive accuracy of 0.05%. This makes it ideal for analyzing the properties of next-generation electronic components, such as power supply inductors, thanks to its maximum measurement frequency of 8 MHz (*1).

Measurement modes  LCR (Measurement with single condition), Continuous testing (Continuous measurement under saved conditions)
Measurement parameters  Z, Y, θ, X, G, B, Q, Rdc (DC resistance), Rs (ESR), Rp, Ls, Lp, Cs, Cp, D (tanδ), σ, ε
 Measurement range  100 mΩ to 100 MΩ, 10 ranges (All parameters are determined according to Z)
 Display range  Z: 0.00 m to 9.99999 GΩ, Y: 0.000 n to 9.99999 GS, θ: ± (0.000° to 180.000°), Q: ± (0.00 to 9999.99), Rdc: ± (0.00 m to 9.99999 GΩ),
D: ± (0.00000 to 9.99999), Δ%: ± (0.000 % to 999.999 %), or other
 Basic accuracy  Z ±0.05% rdg. θ: ±0.03° (representative value, Measurable range: 1 mΩ to 200 MΩ)
 Measurement frequency  4 Hz to 8 MHz (5 digits setting resolution, minimum resolution 10 mHz)
 Measurement signal level  [Normal mode: V mode/CV mode]
4 Hz to 1.0000 MHz: 10 mV to 5 Vrms (maximum 50 mArms)
1.0001 MHz to 8 MHz: 10 mV to 1 Vrms (maximum 10mArms)
[Low impedance high accuracy mode: V mode/CV mode]
4 Hz to 1.0000 MHz: 10 mV to 1 Vrms (maximum 100 mArms)
[Normal mode: CC mode]
4 Hz to 1.0000 MHz: 10 μA to 50 mArms (maximum 5 Vrms)
1.0001 MHz to 8 MHz: 10 μA to 10 mArms (maximum 1 Vrms)
[Low impedance high accuracy mode: CC mode]
4 Hz to 1.0000 MHz: 10 μA to 100 mArms (maximum 1 Vrms)
[DC resistance measurement]
Measurement signal level: Fixed at 1 V
 DC bias measurement  Generating range: DC voltage 0 V to 2.50 V (10 mV resolution)
In low Z high accuracy mode: 0 V to 1 V (10 mV resolution)
 Output impedance  Normal mode: 100 Ω, Low impedance high accuracy mode: 10 Ω
 Display  5.7-inch color TFT with touch panel
 Functions  Comparator, BIN measurement (10 categories for 2 measurement parameters), Trigger function, Open/short compensation, Contact check, Panel loading/saving, Memory function
Interfaces EXT. I/O( HANDLER) ,USB, USB flash drive, LAN, GP-IB, RS-232C, BCD
Power supply  100 to 240 V AC, 50/60 Hz, 50 VA max.
Dimensions and mass  330 mm (12.99 in) W × 119 mm (4.69 in) H × 230 mm (9.06 in) D, 4.2 kg (148.1 oz)
Included accessories  Power cord ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1
Image

Contact Info

Headquarters
963, Jalan Perusahaan, Kawasan Perindustrian Prai, 13600 Prai, Pulau Pinang, Malaysia.
Our Offices
Malaysia | Thailand | Singapore-Indonesia | Vietnam | Philippines

Register Form