Our Products

Product Range

screenshot_2023-11-06_120201
GW Instek GPM-8213 power meter is designed specifically for single-phase (1P/2W) AC power supply’s power measurements. Powerful features, including 4” TFT LCD, five-digit measurement display, 19 power measurement parameters, integral measurement function, high-accuracy voltage/current/power measurem ...Read more
GW Instek - GPM-8213 Digital Power Meter

FEATURES :

  • 4” TFT LCD
  • Two data display modes
  • Standard : two major measurement items + six secondary measurement items
  • Simple : displaying test data of four different measurement items
  • Met the requirement of IEC 62301 power measurement
  • Voltage/ current test frequency bandwidth : DC~6kHz
  • Watt resolution : 1mW
  • Lowest current range : 5mA, resolution : 0.1uA
  • Current/voltage measurements reach CF=3 for distorted wave and CF=6 for half range
  • W-h (power vs time)/A-h (current vs time) integration function
  • Total harmonic distortion measurement



  • Details
  • Specifications

GW Instek GPM-8213 power meter is designed specifically for single-phase (1P/2W) AC power supply’s power measurements. Powerful features, including 4” TFT LCD, five-digit measurement display, 19 power measurement parameters, integral measurement function, high-accuracy voltage/current/power measurement capabilities, front/rear panel input terminals, and various communications ports, are to facilitate users with clear, convenient, and accurate power measurements.

GPM-8213 Specifications

The specifications apply when it warmed up for at least 30 minutes and operates in the slow rate.

Operating Environment: 18~28 °C (64.4~82.4°F)

 

Input

Item

Spec.

Input voltage

600 Vrms

Input current

20 Arms

Input impedance(50/60 Hz)

Voltage

2.4MΩ

Current

5mA - 200mA

500mΩ

0.5A - 20A

5mΩ

Maximum display voltage

700 Vrms

Maximum display current

25 Arms

Maximum allowable isolation voltage

300 V

Low frequency filter

Cut-off frequency

500 Hz

Display

Synchronization frequency

45Hz~ 6kHz

Average

1, 2, 4, 8, 16, 32, 64

Displayed items(Standard mode)

8 items simultaneously.

Displayed items(Simple mode)

4 items simultaneously.

Displayed digits

5

Voltage converter (PT ratio)

1 to 9999.999

Current converter (CT ratio)

1 to 9999.999

Measurement items

Voltage, current, active power, apparent power, reactive power, power factor, phase angle, frequency, integrated current, integrated power, positive integrated power, negative integrated power, integration time, voltage crest factor, current crest factor, voltage peak, current peak, Thd

Displayed measurement parameters

Vdc, Vrms, V+pk, V-pk, Idc, Irms, I+pk, I-pk, P, P+pk, P-pk, VA, VAR, PF, CFV, CFI, DEG, VHz, IHz, THDV, THDI

 Voltage Measurement

Measurement range

CF=3 :

15V, 30V, 60V, 150V, 300V, 600V

CF=6 :

7.5V, 15V, 30V, 75V, 150V, 300V

Crest factor

3, 6

Accuracy

Effective range

1 % to 105 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz f 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz f 1kHz

±(0.1 % reading  +  0.2 % range)

1 kHz f 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

Residual noise

0.5 % of range

 Current Measurement

Measurement range

CF=3 :

5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A, 10A, 20A

CF=6 :

2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A

Crest factor

3, 6

Accuracy

Effective range

1 % to 105 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz f 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz f 1kHz

±(0.1 % reading  +  0.2 % range)

1 kHz f 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

Residual noise

0.5 % of range

 Power Measurement 

Accuracy

Effective range

1 % to 110 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz f 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz f 1kHz

±(0.1 % reading  +  0.3 % range)

1 kHz f 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

 Frequency Measurement 

Measurement range

The filter is turned on

30.000Hz to 499.99Hz

The filter is turned off

30.000Hz to 9.9999kHz

Measurement items

Voltage, Current

Effective input range

10% to 105% of voltage input range

Accuracy

±(0.06 % reading)

Integrator Measurement 

Integrator

Accuracy

±(Accuracy of voltage or current+ 0.1 % reading)

Time

Range

0 hour 0 minute to 9999 hours 59 minutes

Accuracy

±0.01% ±1second

Image

Contact Info

Headquarters
963, Jalan Perusahaan, Kawasan Perindustrian Prai, 13600 Prai, Pulau Pinang, Malaysia.
Our Offices
Malaysia | Thailand | Singapore-Indonesia | Vietnam | Philippines

Register Form

Cancel

Sign in to your account

Register Form

Cancel

Sign in to your account