Our Products

Product Range

screenshot_2023-11-06_110316
Testing is Just a Fingertip Away The GLC-9000, leakage current tester, is used to perform leakage current (or called touch current) tests on medical electrical (IEC 60601-1) and general purpose electric & electronic (IEC 60990) equipment. This tester engages with nine measurement networks (or c ...Read more
GW Instek - GLC-9000 Leakage Current Tester

FEATURES :

  • Touch Panel with Color LCD Display
  • 9 Different Measurement Network to Simulate the Resistance of Human Body
  • 8 Different Types of Leakage current for Medical & General Electrical Equipment
  • Various Leakage Current Measuring ModeDC/AC/AC+DC/AC Peak
  • Upper & Lower Limitation for PASS/FAIL Judgment
  • Meter Mode with SELV/CONV Function
  • 50 Preset Test Conditions Conform to IEC 60990; 30 Sets Memories for Customer Defined
  • Various Standard InterfacesRS232/GPIB/USB Host & Device/EXT I/O



  • Details
  • Specifications

Testing is Just a Fingertip Away

The GLC-9000, leakage current tester, is used to perform leakage current (or called touch current) tests on medical electrical (IEC 60601-1) and general purpose electric & electronic (IEC 60990) equipment. This tester engages with nine measurement networks (or called Measuring Device) to provide the simulation of human body while the EUT (equipment under test) is taking a leakage current testing in compliance with the specific standards or regulations such as IEC, UL, JIS...etc..

In order to provide a simple operation environment, the GLC-9000 equips with large TFT LCD touch panel to configure system as well as to present the measurement settings information and result simultaneously. Besides, there are 50 preset testing conditions, which conform to IEC60990 and other standards, for general electric equipment can be recalled to reduce the setting time. In addition, 30 sets empty memory are available for user defined setting.

A Meter mode is also available for the GLC-9000, it uses the measurement terminal (T1/T2) to measure voltage as the same way of ordinary voltmeter. During the voltage measurement, the SELV function (safety extra low voltage) is applicable to detect whether the voltage value between measuring points exceed the SELV setting.

  • Simple and Intuitive System

  • Various Measurement Network

  • Preset Test Conditions

  • Various Standard Interfaces

  • Measurement Network (MD)

GLC-9000 Specifications

The specifications apply when the GLC-9000 is powered on for at least 30 minutes under +20°C~+30°C.

General

Display

320 x 240 Color TFT LCD

 

Operation

Touch panel

 

Measurement Device

9 kinds

DC

Ranges

50uA/500uA/5mA/25mA

 

Resolution

0.01uA/0.1uA/1uA/10uA

 

Accuracy

±1.0% fs/±(0.2% rdg+3dgt)

AC/AC+DC

Ranges

50uA/500uA/5mA/25mA

 

Resolution

0.01uA/0.1uA/1uA/10uA

 

Accuracy

±2.0% fs/±(2.0% rdg+6dgt) -- 10Hzf100kHz

 

 

±2.0% fs/±(2.0% rdg+10dgt) -- 100kHzf1MHz

AC peak

Ranges

500uA/1mA/10mA/75mA

 

Resolution

0.1uA/1uA/10uA/100uA

 

Accuracy

±4.0%fs/±2.5%fs/±(2.0% rdg+6dgt)
-- 20Hzf1kHz

 

 

±5.0%fs/±5.0%fs/±(5.0% rdg+10dgt)
-- 1kHzf10kHz

Meter Mode

AC/DC/AC+DC

10.0~300.0V

 

AC peak

15.0~430.0V

 

SELV function

Available

EUT (V/I Check)

Voltage range

85V ~ 300V

 

Current range

0.5A ~ 10A

Memory

Empty sets

30 sets

 

Preset sets

50 sets (conform to IEC 60990)

Interfaces

Front panel

USB host

 

Rear panel

RS-232C/GPIB/EXT I/O/USB device

Power Source

For GLC-9000

AC 100V/120V/220V/230V±10%,50/60Hz

 

For EUT

AC 85V~250V,50/60Hz

Dimensions

 

330 (W) x 150 (H) x 350 (D) mm

Weight

 

Approx. 5kg

Image

Contact Info

Headquarters
963, Jalan Perusahaan, Kawasan Perindustrian Prai, 13600 Prai, Pulau Pinang, Malaysia.
Our Offices
Malaysia | Thailand | Singapore-Indonesia | Vietnam | Philippines

Register Form

Cancel

Sign in to your account

Register Form

Cancel

Sign in to your account